Installing and securing the transducer
Lower the transducer to an appropriate depth. With long drops, it may be necessary to use the weighted nose cone (option -WN).
Do not drop the instrument or allow it to free fall down a well as this may damage the sensor.
Secure the body using cable ties or tape. A Heyco cable grip can also be used to secure the transducer to a PVC pipe; see Heyco cable grip for more information. Secure and strain relief the cable using cable ties or the split mesh cable grip. If installing in a well, fasten the cable to the well head.
Never suspend the CS451/CS456 from the connections at the top of the cable. Sharp bends or excessive pinching of the cable can cause damage and may pinch off the vent tube causing measurement errors.
After installation, take several readings to ensure proper operation.