Changing nose cones

The procedure to change or replace a nose cone follows:

  1. Remove the two screws (Screws on the standard nose cone (top), NPT nose cone (middle), and weighted nose cone (bottom)) using a 0.05-inch hex wrench.

  2. Gently pull the nose cone away from the sensor housing.

  3. Insert the new nose cone into the sensor housing.

  4. Replace and tighten the screws.

CAUTION:

Do not use pliers or a wrench on the titanium housing tube. Using these can deform the housing and prevent the nose cone from seating properly.

Screws on the standard nose cone (top), NPT nose cone (middle), and weighted nose cone (bottom)