DataList

Specifies the data to be retrieved from the sensor. If DataList = 1, only concentration and status are returned. If DataList = 2, then laser temperature, pressure, and DC Current are returned in addition to concentration and status. If DataList = 3, then the TGA analog signal 1 and TGA temperatures are also returned. If DataList = 4, then all data except DutyCycle1 and DutyCycle2 are returned. If DataList = 5, then all data are returned. See the list below for complete information:

Name Description DataList ScanMode
ConcA Trace gas concentration measured in ramp A (ppmv) 1 1
ConcB Trace gas concentration measured in ramp B (ppmv) 1 2
ConcC Trace gas concentration measured in ramp C (ppmv) 1 3
TGAStatus Status flags: see table below 1 1
TGAPressure Pressure in the analyzer vacuum manifold (mb) 2 1
LaserTemp emperature of the laser mounting plate (K) 2 1
DCCurrentA Laser DC current for ramp A (mA) 2 1
DCCurrentB Laser DC current for ramp B (mA) 2 2
DCCurrentC Laser DC current for ramp C (mA) 2 3
TGAAnalog1 TGA Analog Channel 1 voltage (V) 3 1
TGATemp1 Temperature of TGA detector mounting plate (°C) 3 1
TGATemp2 Temperature of TGA dewar mounting plate (°C)> 3 1
LaserHeater Voltage applied to the heater on the laser mounting plate to maintain the laser at the specified temperature (V) 4 1
RefDetSignalA Reference detector signal at the center of the spectral scan for ramp A (mV) 4 1
RefDetSignalB Reference detector signal at the center of the spectral scan for ramp B (mV) 4 2
RefDetSignalC Reference detector signal at the center of the spectral scan for ramp C (mV) 4 3
RefDetTransA Reference detector transmittance at the center of the spectral scan for ramp A (%) 4 1
RefDetTransB Reference detector transmittance at the center of the spectral scan for ramp B (%) 4 2
RefDetTransC Reference detector transmittance at the center of the spectral scan for ramp C (%) 4 3
RefDetTemp Reference detector temperature (°C) 4 1
RefDetCoole Current applied to the thermoelectric cooler to maintain the reference detector at its specified temperature (arb) 4 1
RetDetGainOffset Gain and offset settings for the reference detector preamplifier (arb) 4 1
mpDetSignalA Sample detector signal at the center of the spectral scan for ramp A (mV) 4 1
SmpDetSignalB Sample detector signal at the center of the spectral scan for ramp B (mV) 4 2
SmpDetSignalC Sample detector signal at the center of the spectral scan for ramp C (mV) 4 3
SmpDetTransA Sample detector transmittance at the center of the spectral scan for ramp A (%) 4 1
SmpDetTransB Sample detector transmittance at the center of the spectral scan for ramp B (%) 4 2
SmpDetTransC Sample detector transmittance at the center of the spectral scan for ramp C (%) 4 3
SmpDetTemp Sample detector temperature (°C) 4 1
SmpDetCooler Current applied to the thermoelectric cooler to maintain the sample detector at its specified temperature (arb) 4 1
SmpDetGainOffset Gain and offset settings for the sample detector preamplifier (arb) 4 1
DutyCycle1 Duty cycle for the TGA heater 1 5 1
DutyCycle2 Duty cycle for the TGA heaster 2 5 1

Status Flags - The TGAStatus value gives an indication of the overall status of the TGA. A value of zero indicates a normal condition. A nonzero value indicates one or more of the bits are set. The meaning of each of the bits is given below.

Bit

Decimal Value

Description

0 1 Line Lock for ramp A is OFF
1 2 Line Lock for ramp B is OFF
2 4 Line Lock for ramp C is OFF
3 8 Sample detector signal exceeded input range
4 16 Reference detector signal exceeded input range
5 32 Sample detector temperature is outside its specified range
6 64 Reference detector temperature is outside its specified range
7 128 Laser temperature outside its specified range
8 256 Pressure is above is upper limit

Type: Constant