DataList
Specifies the data to be retrieved from the sensor. If DataList = 1, only concentration and status are returned. If DataList = 2, then laser temperature, pressure, and DC Current are returned in addition to concentration and status. If DataList = 3, then the TGA analog signal 1 and TGA temperatures are also returned. If DataList = 4, then all data except DutyCycle1 and DutyCycle2 are returned. If DataList = 5, then all data are returned. See the list below for complete information:
Name | Description | DataList | ScanMode |
---|---|---|---|
ConcA | Trace gas concentration measured in ramp A (ppmv) | 1 | 1 |
ConcB | Trace gas concentration measured in ramp B (ppmv) | 1 | 2 |
ConcC | Trace gas concentration measured in ramp C (ppmv) | 1 | 3 |
TGAStatus | Status flags: see table below | 1 | 1 |
TGAPressure | Pressure in the analyzer vacuum manifold (mb) | 2 | 1 |
LaserTemp | emperature of the laser mounting plate (K) | 2 | 1 |
DCCurrentA | Laser DC current for ramp A (mA) | 2 | 1 |
DCCurrentB | Laser DC current for ramp B (mA) | 2 | 2 |
DCCurrentC | Laser DC current for ramp C (mA) | 2 | 3 |
TGAAnalog1 | TGA Analog Channel 1 voltage (V) | 3 | 1 |
TGATemp1 | Temperature of TGA detector mounting plate (°C) | 3 | 1 |
TGATemp2 | Temperature of TGA dewar mounting plate (°C)> | 3 | 1 |
LaserHeater | Voltage applied to the heater on the laser mounting plate to maintain the laser at the specified temperature (V) | 4 | 1 |
RefDetSignalA | Reference detector signal at the center of the spectral scan for ramp A (mV) | 4 | 1 |
RefDetSignalB | Reference detector signal at the center of the spectral scan for ramp B (mV) | 4 | 2 |
RefDetSignalC | Reference detector signal at the center of the spectral scan for ramp C (mV) | 4 | 3 |
RefDetTransA | Reference detector transmittance at the center of the spectral scan for ramp A (%) | 4 | 1 |
RefDetTransB | Reference detector transmittance at the center of the spectral scan for ramp B (%) | 4 | 2 |
RefDetTransC | Reference detector transmittance at the center of the spectral scan for ramp C (%) | 4 | 3 |
RefDetTemp | Reference detector temperature (°C) | 4 | 1 |
RefDetCoole | Current applied to the thermoelectric cooler to maintain the reference detector at its specified temperature (arb) | 4 | 1 |
RetDetGainOffset | Gain and offset settings for the reference detector preamplifier (arb) | 4 | 1 |
mpDetSignalA | Sample detector signal at the center of the spectral scan for ramp A (mV) | 4 | 1 |
SmpDetSignalB | Sample detector signal at the center of the spectral scan for ramp B (mV) | 4 | 2 |
SmpDetSignalC | Sample detector signal at the center of the spectral scan for ramp C (mV) | 4 | 3 |
SmpDetTransA | Sample detector transmittance at the center of the spectral scan for ramp A (%) | 4 | 1 |
SmpDetTransB | Sample detector transmittance at the center of the spectral scan for ramp B (%) | 4 | 2 |
SmpDetTransC | Sample detector transmittance at the center of the spectral scan for ramp C (%) | 4 | 3 |
SmpDetTemp | Sample detector temperature (°C) | 4 | 1 |
SmpDetCooler | Current applied to the thermoelectric cooler to maintain the sample detector at its specified temperature (arb) | 4 | 1 |
SmpDetGainOffset | Gain and offset settings for the sample detector preamplifier (arb) | 4 | 1 |
DutyCycle1 | Duty cycle for the TGA heater 1 | 5 | 1 |
DutyCycle2 | Duty cycle for the TGA heaster 2 | 5 | 1 |
Status Flags - The TGAStatus value gives an indication of the overall status of the TGA. A value of zero indicates a normal condition. A nonzero value indicates one or more of the bits are set. The meaning of each of the bits is given below.
Bit |
Decimal Value |
Description |
---|---|---|
0 | 1 | Line Lock for ramp A is OFF |
1 | 2 | Line Lock for ramp B is OFF |
2 | 4 | Line Lock for ramp C is OFF |
3 | 8 | Sample detector signal exceeded input range |
4 | 16 | Reference detector signal exceeded input range |
5 | 32 | Sample detector temperature is outside its specified range |
6 | 64 | Reference detector temperature is outside its specified range |
7 | 128 | Laser temperature outside its specified range |
8 | 256 | Pressure is above is upper limit |
Type: Constant