The SampleFieldCal instruction is used to store the values in the FieldCal (Calibrate) file to a data table.
Syntax
DataTable ( TableName, NewFieldCal, Size )
SampleFieldCal
EndTable
'\\\\\\\\\\\\\\\\\\\\ DECLARE VARIABLES //////////////////////////////
Public ZeroMode1,KnownVar1, ZeroMode2,KnownVar2, ZeroMode3,KnownVar3(3)
Public AccelA(1), AccelB(1), AccelC(3)
Units ACCELA = GForce : Units AccelB = GForce : Units AccelC = GForce
Public AccelAmult(1), AccelBmult(1),AccelCmult(3), AccelAoset(1), AccelBoset(1), AccelCoset(3)
Alias AccelA(1) = Accel1 : Alias AccelB(1) = Accel2 : Alias AccelC(1) = Accel3
Alias AccelC(2) = Accel4 : Alias AccelC(3) = Accel5
Public RepA, IndexA, RepB, IndexB, RepC, IndexC
Public LoadTest, Flag(8)
Dim I
DataTable (ACCEL,True,-1) 'Trigger, auto size
DataInterval (0,0,0,0) 'Synchronous, 0 lapses, autosize
Sample (1,AccelA(),IEEE4) '1 Reps,Source,Res
Sample (1,AccelB(),IEEE4) '1 Reps,Source,Res
Sample (3,AccelC(),IEEE4) '3 Reps,Source,Res,Enabled
EndTable
'End of table ACCEL
DataTable (CalTable,NewFieldCal,100) 'Cal Table that stores Calibration values
SampleFieldCal
EndTable
BeginProg 'Program begins here
'MainSequence
'Initialize rep and Index parameters
RepA = 1 : IndexA = 1 : RepB = 1 : IndexB = 1 : RepC = 3 : IndexC = 1
KnownVar2 = 1 ' Set the offset value for ACCELB
AccelAMult(1) = 1 : AccelBMult(1) = 1: AccelAoset(1) = 0 : AccelBoset(1) = 0
For I = 1 To 3 'Do the following to all of BBlk2
AccelCmult(I)=1 : AccelCoset(I) = 0 'Initialize mult & offset values for ACCELC
Next I 'Repeat above until finished
LoadTest = LoadFieldCal(0)
Scan(100,mSec,100,0) 'Scan once every 100 mSec, 100 Scan Buffer, non-burst
' Input Var,Reps,Range,InChan,Excit mV,Reverse,Delay/Settling,Mult Offset
' Setup a two point Calibration function for AccelA
'(Function,Var,Rep,Multiplier, Offset,Mode,KnownVar,Index,Avg)
FieldCal(2,AccelA(),RepA,AccelAmult(1),AccelAoset(1),ZeroMode1,KnownVar1,IndexA,1)
' Setup an offset function for AccelB
'(Function,Var,Rep,Multiplier, Offset,Mode,KnownVar,Index,Avg)
FieldCal (1,AccelB(),RepB,0,AccelBoset(),ZeroMode2,KnownVar2,IndexB,1)
' Setup a zero function for the 3 reps of AccelC
'(Function,Var,Rep,Multiplier, Offset,Mode, KnownVar,Index, Avg)
FieldCal (0,AccelC(),RepC,0,AccelCoset(),ZeroMode3,0,IndexC,1)
CallTable ACCEL
CallTable CalTable
NextScan
EndProg
Remarks
The SampleFieldCal instruction is placed within the data table declaration. The NewFieldCal function can be used to trigger data storage to the table when a new FieldCal file has been written.