Pulse measurement specifications
Switch-closure input
Terminals:
- P_SW
- C1-C2 (Requires an external 100 kΩ resistor connected from the terminal to VX1 or VX2.)
Maximum Input Frequency: 150 Hz
Minimum Switch Closed Time:
Minimum Switch Open Time:
Maximum Bounce Time: 1 ms open without being counted
High-frequency input
Terminals:
- SE1-SE4
- P_LL
- P_SW
- C1-C2
Maximum Input Frequency:
- SE1-SE4: 35 kHz
- P_LL: 20 kHz
- P_SW: 35 kHz
- C1-C2: 3 kHz
Low-level AC input
Terminals: P_LL
Maximum Input Voltage: ±20 VDC
DC-offset Rejection: Internal AC coupling eliminates DC-offset voltages up to ±0.05 VDC
Input Hysteresis: 12 mV at 1 Hz
Low-Level AC Pulse Input Ranges:
Sine wave (mV RMS) | Range (Hz) |
---|---|
20 | 1.0 to 20 |
200 | 0.5 to 200 |
2000 | 0.3 to 10,000 |
5000 | 0.3 to 20,000 |
Quadrature input
Terminals: SE1 and SE2 or C1 and C2 can be configured as digital terminal pairs to monitor the two sensing channels of an encoder.
Maximum Frequency: 2.5 kHz
See Pulse measurements for additional information.