Period-averaging measurement specifications
Use PeriodAvg()
to measure the period (in microseconds) or the frequency (in Hz) of a signal on a single-ended channel.
Terminals: SE1-SE16
Accuracy: ±(0.01% of measurement + resolution), where resolution is 0.13 µs divided by the number of cycles to be measured
Ranges:
- Minimum signal centered around specified period average threshold.
- Maximum signal centered around data logger ground.
- Maximum frequency = 1/(2 * [minimum pulse width]) for 50% duty cycle signals
Gain code option |
Voltage gain |
Minimum peak to peak signal (mV) |
Maximum peak to peak signal (V) |
Minimum pulse width (µs) |
Maximum frequency (kHz) |
---|---|---|---|---|---|
0 | 1 | 500 | 10 | 2.5 | 200 |
1 | 2.5 | 50 | 2 | 10 | 50 |
2 | 12.5 | 10 | 2 | 62 | 8 |
3 | 64 | 2 | 2 | 100 | 5 |
See also Period-averaging measurements.